Mô tả ngắn
Hãng sản xuất: Kruss
The versatile instrument for semi-automatic wettability analysis
The Standard configuration of the DSA30 Drop Shape Analyzer is designed for the semi-automatic measurement of the static and dynamic contact angle and the scientific assessment of wettability by means of the surface free energy (SFE). The instrument is well-suitable for the reliable quality assurance of surface pre-treatment processes as well as the development and optimization of wetting and coating procedures. Accessories for ultrafast one-click SFE measurement as well as wetting analysis at higher temperatures enhance the spectrum of the instrument.
HỖ TRỢ NHANH
THÔNG TIN SẢN PHẨM
- Specifications
- Applications
- Measuring Methods
- Measuring Results
INSTRUMENT DIMENSIONS
Footprint | 610 mm × 250 mm (W × D) |
---|---|
Height | 610 mm |
Weight (without accessories) | 10 kg |
TEMPERATURE MEASUREMENT
Range of temperature measurement | -50 to 400 °C |
---|---|
Resolution | 0.1 °C |
Precision | 0.1 °C |
Accuracy | 1/3 DIN B (±0.1 °C at 0 °C to ±0.8 °C at 400 °C) |
Location of temperature measurement | sample stage chamber cuvette |
External sensor | 2 connectors (PT100) |
POWER SUPPLY
Voltage | 88 to 264 VAC |
---|---|
Power consumption | 100 W |
Frequency | 50 to 60 Hz |
SAMPLE DIMENSIONS
Maximum sample size | 320 mm × ∞ × 275 mm (W × D × H) (without axes) |
---|---|
Maximum measuring surface | – |
ENVIRONMENT
Operating temperature | 10 to 40 °C |
---|---|
Humidity | without condensation |
TEMPERATURE CONTROL
Equipment | temperature-controlled sample stage chambers cuvette |
---|---|
Type of temperature control | liquid electrical Peltier |
Range of temperature control | -30 (with Tempering Chamber – TC40) to 400 °C (with Tempering Chamber – TC21) |
Maximum sample space | 132 mm × 132 mm × 27 mm (W × D × H) (with Tempering Chamber – TC11) |
Resolution | 0.1 K |
Flow-through thermostat | with liquid |
Inert gas | yes |
INTERFACES
PC | USB 3.0 |
---|
HOUSING AND PERIPHERALS
Compartment | – |
---|---|
Needle protection shield | no |
Camera und optics housing | – |
Control keyboard | PC keyboard for ADVANCE software operation available (KB20) |
Levelling | yes |
SOFTWARE
Software type | ADVANCE |
---|---|
Available software modules | contact angle surface free energy of solids interfacial and surface tension of liquids Adhesion Analysis Remote Control API |
SURFACE FREE ENERGY OF SOLIDS
Available | included |
---|---|
Results | surface free energy (SFE) |
Models | equation of state Zisman Fowkes Wu Owens-Wendt-Rabel-Kaelble Schultz-1 extended Fowkes acid-base theory |
CAPTIVE BUBBLE
Available | included |
---|---|
Results | contact angle (CA) |
Range | 0 to 180° (software-based) |
Resolution | 0.01° (software-based) |
Accuracy | 0.1° (instrument-based) |
Models | conic section polynom circle Young-Laplace height-width |
Type | advancing receding static dynamic tilting (additional accessories may be required) |
SESSILE DROP
Available | included |
---|---|
Results | contact angle (CA) |
Range | 0 to 180° (software-based) |
Resolution | 0.01° (software-based) |
Accuracy | 0.1° (instrument-based) |
Models | conic section polynom circle Young-Laplace height-width |
Type | advancing receding static dynamic tilting (additional accessories may be required) |
PENDANT DROP/RISING DROP
Available | optional |
---|---|
Results | interfacial tension (IFT)/surface tension (SFT) |
Range | 0.01 to 2000 mN/m (software-based) |
Resolution | 0.01 mN/m (software-based) |
Models | Young-Laplace |
Types (Pendant drop/Rising drop) | static, dynamic |
CAMERA SYSTEM (STANDARD)
Type | Camera CF04 |
---|---|
Connection | USB 3.0 |
Resolution | 1920 × 1200 px |
Frame rate | 2300 fps |
Dark noise | 7 electrons |
Dynamic range | 73 dB |
ILLUMINATION
Type | high power monochromatic LED |
---|---|
Wave length, dominant | 470 nm |
Field of light | Ø 42 mm |
OPTICS (STANDARD)
Focus | manual |
---|---|
Zoom | 6.5× zoom, manual |
View angle | ±3° |
Field of view | CF04: 3.2 mm × 3.2 mm to 18.5 mm × 18.5 mm |
Resolution of optics | CF04: 2.5 to 16.2 μm |
CAMERA SYSTEM (OPTIONAL)
Type | High Speed Camera CF06 |
---|---|
Connection | USB 3.0 |
Resolution | 640 × 480 px |
Frame rate | 3400 fps |
Dark noise | 10.5 electrons |
Dynamic range | 56.6 dB |
DOSING SYSTEM
Dosing | software-controlled |
---|---|
Drop deposition | software-controlled |
Cartridge/syringe (volume) | glass (450 μL), disposable (900 μL) |
Resolution | 0.1 μL |
Speed | 0.02 to 25 µL/s |
STAGES
Control | x-axis: manual (software-controlled optional available) y-axis: manual (software-controlled optional available) z-axis: manual |
---|---|
Range | – |
Length | x-axis: 100 mm y-axis: 100 mm z-axis: 45 mm |
Resolution | x-axis: 2 mm/turn y-axis: 2 mm/turn z-axis: 16 mm/turn |
Accuracy | x-axis: – y-axis: – z-axis: – |
OPTICS (OPTIONAL)
Focus | manual |
---|---|
Zoom | 6.5× zoom, manual |
View angle | ±3° |
Field of view | CF06: 1.4 mm × 1 mm to 8.1 mm × 6 mm |
Resolution of optics | CF06: 2.1 to 13.3 μm |
DOUBLE PRESSURE DOSING SYSTEM (OPTIONAL)
Drop deposition | software-controlled |
---|---|
Cartridge, volume | disposable (1 mL) |
Resolution | 0.1 μL |
Speed | fixed |
TILTING (OPTIONAL)
Types | internal tilting table |
---|---|
Control | software-controlled |
Range | 0 to 90° |
Resolution | 0.01° |
Accuracy | 0.3° |
DOUBLE PRESSURE DOSING SYSTEM
Drop deposition | – |
---|---|
Cartridge, volume | – |
Resolution | – |
Speed | – |
- Quality assurance and optimization of activation processes such as plasma treatment, flame treatment, or corona treatment before coating or bonding
- Analyzing the wettability of plastics, glass, ceramics, wood, paper, or metal
- Analyzing surface cleanliness
- Wetting analysis at controlled temperature
- Contact angle of a drop on a solid surface
- Contact angle using a gas bubble beneath a solid surface in a liquid
- Contact angles of two drops dosed in parallel
- Surface free energy of a solid using contact angle data
- Surface tension using the curvature of a perfectly symmetrical drop on a circular sample pedestal
- Roll-off behavior and advancing/receding contact angle of a drop on a tilted surface
- Carrying out a measurement up to 20 consecutive times and displaying the results together
- Remote control of ADVANCE for integration in custom-designed, complex automations
- Static contact angle
- Advancing and receding contact angle
- Contact angle using a bubble in a liquid
- Surface free energy (SFE) according to the following models: Owens-Wendt-Rabel-Kaelble (OWRK), Fowkes, Wu, Zisman, Equation of State
- Temperature