Hãng sản xuất: Kruss
Automatic wettability analysis of large surfaces
The Large version of the DSA100 Drop Shape Analyzer is prepared for automated wettability analysis by means of the contact angle or even the surface free energy. Fitted in a particularly wide frame and equipped with a large, rotating sample table plus an extra-long movement axis, the instrument can handle samples of an area of up to 500 × 500 mm.
HỖ TRỢ NHANH
THÔNG TIN SẢN PHẨM
- Specifications
- Applications
- Measuring methods
INSTRUMENT DIMENSIONS
Footprint | 1000 mm × 375 mm (W × D) |
---|---|
Height | 490 mm |
Weight (without accessories) | 34 kg |
TEMPERATURE MEASUREMENT
Range of temperature measurement | -50 to 400 °C |
---|---|
Resolution | 0.1 °C |
Precision | 0.1 °C |
Accuracy | 1/3 DIN B (±0.1 °C at 0 °C to ±0.8 °C at 400 °C) |
Location of temperature measurement | environment air |
External sensor | 2 connectors (PT100) |
POWER SUPPLY
Voltage | 88 to 264 VAC |
---|---|
Power consumption | 100 W |
Frequency | 50 to 60 Hz |
SAMPLE DIMENSIONS
Maximum sample size | 700 mm × ∞ × 275 mm (W × D × H) (without axes) |
---|---|
Maximum measuring surface | 500 mm × 500 mm (W × D) |
ENVIRONMENT
Operating temperature | 10 to 40 °C |
---|---|
Humidity | without condensation |
TEMPERATURE CONTROL
Equipment | – |
---|---|
Type of temperature control | – |
Range of temperature control | – |
Maximum sample space | – |
Resolution | – |
Flow-through thermostat | – |
Inert gas | – |
INTERFACES
PC | USB 3.0 |
---|
HOUSING AND PERIPHERALS
Compartment | test liquids protected against light |
---|---|
Needle protection shield | yes |
Camera und optics housing | yes |
Control keyboard | PC keyboard for ADVANCE software operation available (KB20) |
Levelling | yes |
SOFTWARE
Software type | ADVANCE |
---|---|
Available software modules | contact angle surface free energy of solids interfacial and surface tension of liquids Adhesion Analysis Remote Control API |
SURFACE FREE ENERGY OF SOLIDS
Available | included |
---|---|
Results | surface free energy (SFE) |
Models | equation of state Zisman Fowkes Wu Owens-Wendt-Rabel-Kaelble extended Fowkes acid-base theory |
CAPTIVE BUBBLE
Available | included |
---|---|
Results | contact angle (CA) |
Range | 0 to 180° (software-based) |
Resolution | 0.01° (software-based) |
Accuracy | 0.1° (instrument-based) |
Models | conic section polynom circle Young-Laplace height-width |
Type | advancing receding static dynamic tilting (additional accessories may be required) |
SESSILE DROP
Available | included |
---|---|
Results | contact angle (CA) |
Range | 0 to 180° (software-based) |
Resolution | 0.01° (software-based) |
Accuracy | 0.1° (instrument-based) |
Models | conic section polynom circle Young-Laplace height-width |
Type | advancing receding static dynamic tilting (additional accessories may be required) |
PENDANT DROP/RISING DROP
Available | included |
---|---|
Results | interfacial tension (IFT)/surface tension (SFT) |
Range | 0.01 to 2000 mN/m (software-based) |
Resolution | 0.01 mN/m (software-based) |
Models | Young-Laplace |
Types (Pendant drop/Rising drop) | static, dynamic |
CAMERA SYSTEM (STANDARD)
Type | Camera CF04 |
---|---|
Connection | USB 3.0 |
Resolution | 1920 × 1200 px |
Frame rate | 2300 fps |
Dark noise | 7 electrons |
Dynamic range | 73 dB |
ILLUMINATION
Type | high power monochromatic LED |
---|---|
Wave length, dominant | 470 nm |
Field of light | 46 mm × 46 mm (D × H) |
OPTICS (STANDARD)
Focus | manual |
---|---|
Zoom | 7× zoom, manual |
View angle | ±4° |
Field of view | CF04: 3.9 mm × 3.9 mm to 24.7 mm × 24.7 mm |
Resolution of optics | CF04: 3.1 to 21.7 μm |
CAMERA SYSTEM (OPTIONAL)
Type | High Speed Camera CF06 |
---|---|
Connection | USB 3.0 |
Resolution | 640 × 480 px |
Frame rate | 3400 fps |
Dark noise | 10.5 electrons |
Dynamic range | 56.6 dB |
DOSING SYSTEM
Dosing | software-controlled |
---|---|
Drop deposition | software-controlled |
Cartridge/syringe (volume) | glass (450 μL), disposable (900 μL) |
Resolution | 0.1 μL |
Speed | 0.02 to 25 µL/s |
STAGES
Control | y-axis: software-controlled z-axis: manual z-axis (optional): software-controlled rotation axis: software-controlled |
---|---|
Range | y-axis: 350 mm z-axis manual: 45 mm z-axis software-controlled (optional): 38 mm rotation axis: 360° |
Length | – |
Resolution | y-axis: 10 μm z-axis manual: 16 mm/turn z-axis software-controlled (optional): 10 μm rotation axis: 0.1° |
Accuracy | y-axis: 100 μm z-axis manual: – z-axis software-controlled (optional): 100 μm rotation axis: 1° |
OPTICS (OPTIONAL)
Focus | manual |
---|---|
Zoom | 7× zoom, manual |
View angle | ±4° |
Field of view | CF06: 1.7 mm × 1.3 mm to 10.8 mm × 8.1 mm |
Resolution of optics | CF06: 2.5 to 17.8 μm |
DOUBLE PRESSURE DOSING SYSTEM (OPTIONAL)
Drop deposition | software-controlled |
---|---|
Cartridge, volume | disposable (1 mL) |
Resolution | 0.1 μL |
Speed | fixed |
TILTING (OPTIONAL)
Types | – |
---|---|
Control | – |
Range | – |
Resolution | – |
Accuracy | – |
DOUBLE PRESSURE DOSING SYSTEM
Drop deposition | – |
---|---|
Cartridge, volume | – |
Resolution | – |
Speed | – |
- Characterization of large pre-treated or coated surfaces
- Investigation of the cleanliness of windows and frames
- Surface analysis for monitors and touch panels
- Characterization of automobile parts before coating or bonding
- Hydrophobicity analysis of ceramic tiles
- Contact angle of a drop on a solid surface
- Automated position-dependent contact angle measurement and evaluation
- Surface free energy of a solid using contact angle data
- Carrying out a measurement up to 20 consecutive times and displaying the results together
- Contact angles of two drops dosed in parallel
- Contact angle using a gas bubble beneath a solid surface in a liquid
- Surface tension using the curvature of a perfectly symmetrical drop on a circular sample pedestal
- Remote control of ADVANCE for integration in custom-designed, complex automations