Mô tả ngắn
Hãng sản xuất: Kruss
The semi-automatic premium solution for wettability analysis
The Standard configuration of the DSA100 Drop Shape Analyzer system solution is designed for the semi-automatic measurement of the static and dynamic contact angle and the scientific assessment of wettability by means of the surface free energy (SFE).
The instrument, which is equipped with a robust housing, is well-suited for the reliable quality assurance of surface pre-treatment processes as well as the development and optimization of wetting and coating procedures. Accessories for ultrafast one-click SFE measurement as well as wetting analysis at higher temperatures enhance the spectrum of the instrument.
HỖ TRỢ NHANH
THÔNG TIN SẢN PHẨM
- Specifications
- Applications
- Measuring Methods
- Measuring Results
INSTRUMENT DIMENSIONS
| Footprint | 555 mm × 375 mm (W × D) |
|---|---|
| Height | 490 mm |
| Weight (without accessories) | 24 kg |
TEMPERATURE MEASUREMENT
| Range of temperature measurement | -50 to 400 °C |
|---|---|
| Resolution | 0.1 °C |
| Precision | 0.1 °C |
| Accuracy | 1/3 DIN B (±0.1 °C at 0 °C to ±0.8 °C at 400 °C) |
| Location of temperature measurement | sample stage chamber cuvette |
| External sensor | 2 connectors (PT100) |
POWER SUPPLY
| Voltage | 88 to 264 VAC |
|---|---|
| Power consumption | 100 W |
| Frequency | 50 to 60 Hz |
SAMPLE DIMENSIONS
| Maximum sample size | 320 mm × ∞ × 275 mm (W × D × H) (without axes) |
|---|---|
| Maximum measuring surface | – |
ENVIRONMENT
| Operating temperature | 10 to 40 °C |
|---|---|
| Humidity | without condensation |
TEMPERATURE CONTROL
| Equipment | temperature-controlled sample stage chamber cuvette |
|---|---|
| Type of temperature control | liquid electrical Peltier |
| Range of temperature control | -30 (with Tempering Chamber – TC40) to 400 °C (with Tempering Chamber – TC21) |
| Maximum sample space | 132 mm × 132 mm × 27 mm (W × D × H) (with Tempering Chamber – TC11) |
| Resolution | 0.1 K |
| Flow-through thermostat | with liquid |
| Inert gas | yes |
INTERFACES
| PC | USB 3.0 |
|---|
HOUSING AND PERIPHERALS
| Compartment | test liquids protected against light |
|---|---|
| Needle protection shield | yes |
| Camera und optics housing | yes |
| Control keyboard | PC keyboard for ADVANCE software operation available (KB20) |
| Levelling | yes |
SOFTWARE
| Software type | ADVANCE |
|---|---|
| Available software modules | contact angle surface free energy of solids interfacial and surface tension of liquids Adhesion Analysis Remote Control API |
SURFACE FREE ENERGY OF SOLIDS
| Available | included |
|---|---|
| Results | surface free energy (SFE) |
| Models | equation of state Zisman Fowkes Wu Owens-Wendt-Rabel-Kaelble Schultz-1 extended Fowkes acid-base theory |
CAPTIVE BUBBLE
| Available | included |
|---|---|
| Results | contact angle (CA) |
| Range | 0 to 180° (software-based) |
| Resolution | 0.01° (software-based) |
| Accuracy | 0.1° (instrument-based) |
| Models | conic section polynom circle Young-Laplace height-width |
| Type | advancing receding static dynamic tilting (additional accessories may be required) |
SESSILE DROP
| Available | included |
|---|---|
| Results | contact angle (CA) |
| Range | 0 to 180° (software-based) |
| Resolution | 0.01° (software-based) |
| Accuracy | 0.1° (instrument-based) |
| Models | conic section polynom circle Young-Laplace height-width |
| Type | advancing receding static dynamic tilting (additional accessories may be required) |
PENDANT DROP/RISING DROP
| Available | optional |
|---|---|
| Results | interfacial tension (IFT)/surface tension (SFT) |
| Range | 0.01 to 2000 mN/m (software-based) |
| Resolution | 0.01 mN/m (software-based) |
| Models | Young-Laplace |
| Types (Pendant drop/Rising drop) | static, dynamic |
CAMERA SYSTEM (STANDARD)
| Type | Camera CF04 |
|---|---|
| Connection | USB 3.0 |
| Resolution | 1920 × 1200 px |
| Frame rate | 2300 fps |
| Dark noise | 7 electrons |
| Dynamic range | 73 dB |
ILLUMINATION
| Type | high power monochromatic LED |
|---|---|
| Wave length, dominant | 470 nm |
| Field of light | 46 mm × 46 mm (D × H) |
OPTICS (STANDARD)
| Focus | manual |
|---|---|
| Zoom | 7× zoom, manual |
| View angle | ±4° |
| Field of view | CF04: 3.9 mm × 3.9 mm to 24.7 mm × 24.7 mm |
| Resolution of optics | CF04: 3.1 to 21.7 μm |
CAMERA SYSTEM (OPTIONAL)
| Type | High Speed Camera CF06 |
|---|---|
| Connection | USB 3.0 |
| Resolution | 640 × 480 px |
| Frame rate | 3400 fps |
| Dark noise | 10.5 electrons |
| Dynamic range | 56.6 dB |
DOSING SYSTEM
| Dosing | software-controlled |
|---|---|
| Drop deposition | software-controlled |
| Cartridge/syringe (volume) | glass (450 μL), disposable (900 μL) |
| Resolution | 0.1 μL |
| Speed | 0.02 to 25 µL/s |
STAGES
| Control | x-axis: manual (software-controlled optional available) y-axis: manual (software-controlled optional available) z-axis: manual |
|---|---|
| Range | – |
| Length | x-axis: 100 mm y-axis: 100 mm z-axis: 45 mm |
| Resolution | x-axis: 2 mm/turn y-axis: 2 mm/turn z-axis: 16 mm/turn |
| Accuracy | x-axis: – y-axis: – z-axis: – |
OPTICS (OPTIONAL)
| Focus | manual |
|---|---|
| Zoom | 7× zoom, manual |
| View angle | ±4° |
| Field of view | CF06: 1.7 mm × 1.3 mm to 10.8 mm × 8.1 mm |
| Resolution of optics | CF06: 2.5 to 17.8 μm |
DOUBLE PRESSURE DOSING SYSTEM (OPTIONAL)
| Drop deposition | software-controlled |
|---|---|
| Cartridge, volume | disposable (1 mL) |
| Resolution | 0.1 μL |
| Speed | fixed |
TILTING (OPTIONAL)
| Types | internal tilting table external tilting table |
|---|---|
| Control | software-controlled |
| Range | 0 to 90° |
| Resolution | internal: 0.01° external: 0.1° |
| Accuracy | internal: 0.3° external: 1° |
DOUBLE PRESSURE DOSING SYSTEM
| Drop deposition | – |
|---|---|
| Cartridge, volume | – |
| Resolution | – |
| Speed | – |
- Quality assurance and optimization of activation processes such as plasma treatment, flame treatment, or corona treatment before coating or bonding
- Analyzing wettability of plastics, glass, ceramics, wood, paper, or metal
- Analyzing the effectiveness of hydrophobic coatings and other coatings
- Analyzing surface cleanliness
- Wetting analysis at controlled temperature
- Contact angle of a drop on a solid surface
- Contact angle using a gas bubble beneath a solid surface in a liquid
- Contact angles of two drops dosed in parallel
- Surface free energy of a solid using contact angle data
- Roll-off behavior and advancing/receding contact angle of a drop on a tilted surface
- Surface tension using the curvature of a perfectly symmetrical drop on a circular sample pedestal
- Carrying out a measurement up to 20 consecutive times and displaying the results together
- Remote control of ADVANCE for integration in custom-designed, complex automations
- Static contact angle
- Advancing and receding contact angle
- Contact angle using a bubble in a liquid
- Surface free energy (SFE) according to the following models: Owens-Wendt-Rabel-Kaelble (OWRK), Fowkes, Wu, Zisman, Equation of State
- Temperature